Short answer: most air quality sensors fall into five types. Optical (light-scattering) sensors measure PM2.5 and PM10. NDIR sensors measure CO2. Electrochemical cells measure gases such as NO2, ozone, CO and SO2. Metal oxide (MOS) sensors and photoionisation detectors (PID) give a total VOC signal. Low-cost versions are useful, but humidity, temperature and drift affect them. In Indian conditions, a PM2.5 sensor needs local calibration against a reference monitor before its numbers can be trusted.
This guide explains how each sensor type works, how accurate they really are, what the US EPA's test targets say, how to calibrate, and what to ask a supplier.
What are the main types of air quality sensor?
The World Meteorological Organization's review of low-cost sensors groups them by how they detect a pollutant (WMO, Low-cost sensors for the measurement of atmospheric composition, 2018):
| Sensor type | Measures | How it works | Main weak points |
|---|---|---|---|
| Optical light scattering (nephelometer or optical particle counter) | PM1, PM2.5, PM10 | A laser or LED lights the air; scattered light is converted to a particle count or mass estimate | Doesn't weigh particles; misreads at high humidity; struggles with very small particles |
| NDIR (non-dispersive infrared) | CO2 | CO2 absorbs infrared light at a known wavelength; less light reaching the detector means more CO2 | Needs periodic checks against a reference, like any sensor |
| Electrochemical (EC) | NO2, O3, CO, SO2, NO | Target gas reacts at an electrode and creates a small current | Cross-sensitivity to other gases; affected by humidity and temperature; limited life |
| Metal oxide semiconductor (MOS) | VOCs, CO, some others | Gas sticks to a heated film and changes its resistance | Non-linear; reacts to many gases; sensitive to the environment |
| Photoionisation detector (PID) | Total VOCs | UV lamp ionises organic molecules and the current is measured | Reads different VOCs with different efficiency, so totals depend on the mix |
How accurate are low-cost PM2.5 sensors?
Optical PM sensors don't weigh particles. They count and size them by light, then assume a density and size distribution to estimate mass. The WMO review points out several limits:
- Low-cost light-scattering devices see particles in roughly the 400 to 10,000 nm range for number counts. Lower-cost units detect only particles larger than about 300 nm.
- Reported detection limits are around 1 to 10 µg/m³, usually under good lab conditions.
- Upper detection limits are typically 500 to 1,000 µg/m³, so they're "unsuitable for extremely polluted locations" without checking.
- Water is the biggest problem. Performance degrades when relative humidity goes above about 80 to 85%, because particles absorb water and grow.
That last point matters in India. Monsoon months, coastal cities and winter fog all push humidity past that range.
What does real Indian data show?
The best evidence is a long-term study that ran PurpleAir PA-II sensors next to reference beta attenuation monitors in Delhi, rural Hamirpur in Uttar Pradesh and Bengaluru (Campmier et al., Atmospheric Measurement Techniques, 2023).
Without calibration, the sensors tracked the reference reasonably (R² of 0.55 to 0.74) but were inaccurate, with normalised error of 40% or more. With a simple annual calibration model, error fell to 20 to 30% and R² rose to 0.82 to 0.95. Seasonal calibration did better still.
The authors found the best calibration model was different at each site, because aerosol mix and weather vary across India. Their advice: calibrate locally where you can. A correction built in California or Europe won't transfer cleanly to a Delhi winter.
What performance should you ask for?
The US EPA published performance targets for PM2.5 sensors used in outdoor, non-regulatory monitoring in February 2021 (US EPA, EPA/600/R-20/280). It later added targets for PM10, NO2, CO and SO2 in February 2024 (US EPA Air Sensor Toolbox).
| Metric (PM2.5, base field testing) | EPA target |
|---|---|
| Precision | Standard deviation ≤ 5 µg/m³ or coefficient of variation ≤ 30% |
| Bias: slope | 1.0 ± 0.35 |
| Bias: intercept | Between -5 and +5 µg/m³ |
| Linearity: R² | ≥ 0.70 |
| Error | RMSE ≤ 7 µg/m³ or NRMSE ≤ 30% |
Base testing means at least three sensors beside a reference monitor for at least 30 days at each of two sites in different climates. EPA also suggests lab tests at 40% and 85% relative humidity, at 20°C and 40°C, and at high concentrations of 150 and 250 µg/m³. Ask your supplier for results against these numbers, ideally from an Indian site.
These are targets for supplementary monitoring, not regulatory compliance. For CPCB-compliant ambient monitoring you need reference-grade analysers. Our CPCB ambient monitoring checklist covers that.
Which sensor suits which job?
| Use | Sensor approach | Why |
|---|---|---|
| Office or school indoor air | Optical PM, NDIR CO2, MOS or PID VOC, temperature and humidity | CO2 shows ventilation; PM shows outdoor ingress |
| Plant boundary or construction dust | Optical PM with heated inlet or humidity correction, plus wind | Fast trend data and alerts between reference checks |
| City or campus network | Many calibrated sensor nodes plus one reference station | Reference station anchors the network calibration |
| Stack emissions | Not low-cost sensors: use CEMS | Regulatory method required |
| Ambient compliance (CAAQMS) | Reference analysers | Required by CPCB |
For indoor air in offices, see our indoor air quality guide and IAQ monitoring products. For dust, see particulate monitoring.
How do you calibrate a sensor network?
- Collocate first. Run a few sensors beside a reference monitor (a CPCB station or your own BAM) for several weeks.
- Build a correction. Typically a regression using the raw sensor value, relative humidity and temperature.
- Check by season. Re-collocate when the season changes. The Indian study found seasonal models beat a single annual one.
- Rotate units. Move a "travelling" sensor between the reference site and the field to catch drift.
- Version your corrections. Keep old raw data so you can re-process it when the model changes.
The WMO review adds a fair warning: long-term stability of low-cost PM sensors beyond a year is uncertain, and most evaluations last only weeks or months. Plan for replacement and re-checks.
Do sensors need any approval in India?
Low-cost sensors used for information, not compliance, don't need CPCB approval. If a device connects over a telecom network, check whether it falls under the Telecommunication Engineering Centre's standards. On 19 June 2026, TEC notified revised Essential Requirements for several device types, including "End Point Device for Environmental Monitoring" and IoT gateways. Old and new versions run together for 90 days, after which only the revised standard applies (Gazette of India, S.O. 3349(E)). Ask your supplier how their device meets it.
Questions to ask an air quality sensor supplier
- Which sensing principle does each channel use, and who makes the sensor element?
- Do you have collocation results against a reference monitor in India? Share R², slope, intercept and RMSE.
- How do you handle relative humidity above 80%?
- What is the upper measurement limit for PM2.5, and what happens above it?
- How often should the sensor be recalibrated or replaced, and what does that cost?
- Do you keep raw data alongside corrected data?
- Does the device meet the TEC requirements for environmental monitoring end-point devices?
FAQs
What is the best sensor for PM2.5? For low-cost use, a laser optical particle sensor with humidity correction and local calibration. For compliance, a reference method such as a beta attenuation monitor.
Which sensor measures CO2 accurately? NDIR is the usual low-cost method for CO2. Like any sensor, check it against a reference from time to time.
Why do PM sensors read high during fog or monsoon? Particles absorb water and grow at high humidity, so the optical sensor sees bigger particles. Performance degrades above about 80 to 85% relative humidity.
Can low-cost sensors replace a CAAQMS station? No. They can fill gaps between reference stations and give early warnings, but compliance monitoring needs reference analysers.
How often should air quality sensors be calibrated? Check against a reference at least each season in India, and keep a travelling unit to catch drift.
Planning a sensor network for a campus, plant or city? Talk to our environmental team or explore our environmental intelligence solutions.
Sources
- WMO, Low-cost sensors for the measurement of atmospheric composition, 2018: report
- Campmier et al., Seasonally optimized calibrations improve low-cost sensor performance, Atmos. Meas. Tech. 16, 4357-4374, 2023: paper
- US EPA, Performance testing protocols, metrics and target values for PM2.5 air sensors, EPA/600/R-20/280, Feb 2021: record
- US EPA, Air sensor performance targets and testing protocols: Air Sensor Toolbox
- Gazette of India, S.O. 3349(E), TEC revised essential requirements, 19 Jun 2026: egazette.gov.in